Product Details
Properties
| Assay type | Competitive-ELISA |
| Format | 96T/48T |
| Assay time | 2.0h |
| Detection range | 62.50-4000 pg/mL |
| Sensitivity | 37.50 pg/mL |
| Sample type &Sample volume | serum, plasma and other biological fluids; 50?L |
| Specificity | This kit recognizes SOD1 in samples. No significant cross-reactivity or interference between SOD1 and analogues was observed. |
| Reproducibility | Both intra-CV and inter-CV are < 10%. |
| Application | This ELISA kit applies to the in vitro quantitative determination of SOD1 concentrations in serum, plasma and other biological fluids. |
Test Principle
Kit components & Storage
| Item | Specifications | Storage |
|---|---|---|
| Micro ELISA Plate(Dismountable) | 96T: 8 wells ?12 strips 48T: 8 wells ?6 strips |
-20?, 6 months |
| Reference Standard | 96T: 2 vials 48T: 1 vial |
|
| Concentrated Biotinylated Detection Ab (100?) | 96T: 1 vial, 120 ?L 48T: 1 vial, 60 ?L |
|
| Concentrated HRP Conjugate (100?) | 96T: 1 vial, 120 ?L 48T: 1 vial, 60 ?L |
-20?(Protect from light), 6 months |
| Reference Standard & Sample Diluent | 1 vial, 20 mL | 2-8?C, 6 months |
| Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
| HRP Conjugate Diluent | 1 vial, 14 mL | |
| Concentrated Wash Buffer (25?) | 1 vial, 30 mL | |
| Substrate Reagent | 1 vial, 10 mL | 2-8?(Protect from light) |
| Stop Solution | 1 vial, 10 mL | 2-8? |
| Plate Sealer | 5 pieces | |
| Manual | 1 copy | |
| Certificate of Analysis | 1 copy |
Technical Data
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level SOD1 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level SOD1 were tested on 3 different plates, 20 replicates in each plate.
| Intra-assay Precision | Inter-assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean (pg/mL) | 209.30 | 367.40 | 1835.40 | 196.20 | 331.30 | 1975.00 |
| Standard deviation | 13.20 | 14.70 | 55.10 | 12.80 | 16.60 | 75.10 |
| CV (%) | 6.31 | 4.00 | 3.00 | 6.52 | 5.01 | 3.80 |
Recovery
The recovery of SOD1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
| Sample Type | Range (%) | Average Recovery (%) |
|---|---|---|
| Serum(n=8) | 96-108 | 102 |
| EDTA plasma(n=8) | 88-101 | 94 |
| Cell culture media(n=8) | 91-105 | 97 |
Linearity
Samples were spiked with high concentrations of Human SOD1 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Target Information
| Database Links | SwissProt:?P00441 |
| Synonyms | ALS, ALS1, IPOA, Superoxide dismutase [Cu-Zn] |
| Research Area | Cancer,?Cell Biology,?Metabolism,?Signal transduction |
Assay Procedures
![]() |
1. Add 50?L standard or sample to the wells, immediately add 50?L Biotinylated Detection Ab working solution to each well. Incubate for 45 min at 37?C |
![]() |
2. Aspirate and wash the plate for 3 times |
![]() |
3. Add 100?L HRP conjugate working solution. Incubate for 30 min at 37?C. Aspirate and wash the plate for 5 times |
![]() |
4. Add 90?L Substrate Reagent. Incubate for 15 min at 37?C |
![]() |
5. Add 50?L Stop Solution |
![]() |
6. Read the plate at 450nm immediately. Calculation of the results |