VENTICELL 404
Ovens with Forced Circulation
Type Eco Line MC000214 Type Evo Line MC001818 Due to a patented ventilation system the air within the VENTICELL® chamber is ventilated in a regular spiral way. This leads to a homogenous temperature profile throughout the chamber and short heating times. Operating economy is ensured by higher rate and precision of heating in laboratories. Especially suitable for very moist goods.- precise and fast laboratory drying oven
- chamber volume 404 litres
- temperature range from 10°C above ambient to 250/300°C
- forced air circulation system in the chamber guarantees homogeneous temperature distribution in all the processes of drying, heating and sterilization of materials in laboratories
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- higher speed and accuracy of all the tempering procedures ensures economical operation, especially suitable for high humidity materials, for demanding and accurate tests and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Specification
Interior of stainless steel material No. 1.4301
volume | 404 l |
depth | 520 mm |
width | 540 mm |
height | 1410 mm |
Tray
standard equipment | 2 pcs |
storage area (w × d) | 520 × 485 mm |
racks | 19 max. number |
distance between guide rails | 70 mm |
Maximal weight of the load *)
inside the oven | 100 kg/oven |
per tray | 30 kg/tray |
Number of doors
1 pcs |
External dimensions (including door and handle)
depth | 790 mm |
diameter of the air branch outer/inner | 52/49 mm |
width | 760 mm |
height (inclusive foots – N) | 1910 (K) mm |
Package dimensions (three layers carton)
depth | 860 mm |
width | 830 mm |
depth | 2085 mm |
Weight
brut | 175 kg |
net | 150 kg |
Working temperature **) (beginning of the regulation) – VENTICELL / VENTICELL plus
from 10 °C above ambient temp. to | 250 °C |
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door, > 50 °C
time variation cca (±) | 0,4 °C |
space deviation (±) % cca (±)% of the reached temperature | 1,5 % |
Time required to reach 250 °C with closed air flap and voltage 230 V
58 min |
Heat emission at 250 °C
1940 W |
Air exchange speed at 250 °C
18 cca/hour |
Electricity – mains 50/60 Hz
max. power input | 3,7 kW |
nominal voltage | 400/3N V |
current | 5,6 A |
stand by | 5 W |