VENTICELL 111
Ovens with Forced Circulation
-Type Eco Line MC000212
-Type Evo Line MC001816
Due to a?patented ventilation system the air within the VENTICELL??chamber is ventilated in a?regular spiral way. This?leads to a?homogenous temperature profile throughout the?chamber and short heating times. Operating economy is ensured by higher rate and precision of heating in laboratories. Especially suitable for?very moist goods.
- precise and fast laboratory drying oven
- chamber volume 111 litres
- temperature range from 10?C above ambient to 250/300?C
- forced air circulation system in the chamber guarantees homogeneous temperature distribution in all the processes of drying, heating and sterilization of materials in laboratories
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- higher speed and accuracy of all the tempering procedures ensures economical operation, especially suitable for high humidity materials, for demanding and accurate tests and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Specification
Interior of stainless steel material No. 1.4301
| volume | 111 l |
| depth | 370 mm |
| width | 540 mm |
| height | 530 mm |
Tray
| standard equipment | 2 pcs |
| storage area (w ? d) | 520 ? 335 mm |
| racks | 7 max. number |
| distance between guide rails | 70 mm |
Maximal weight of the load *)
| inside the oven | 50 kg/oven |
| per tray | 20 kg/tray |
Number of doors
| 1 pcs |
External dimensions (including door and handle)
| depth | 640 mm |
| diameter of the air branch outer/inner | 52/49 mm |
| width | 760 mm |
| height (inclusive foots – N) | 860 (N) mm |
Package dimensions (three layers carton)
| depth | 730 mm |
| width | 830 mm |
| depth | 1060 mm |
Weight
| brut | 87 kg |
| net | 75 kg |
Working temperature **) (beginning of the regulation) – VENTICELL / VENTICELL plus
| from 10 ?C above ambient temp. to | 250 ?C |
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door, > 50 ?C
| time variation cca (?) | 0,4 ?C |
| space deviation (?) % cca (?)% of the reached temperature | 1 % |
Time required to reach 250 ?C with closed air flap and voltage 230 V
| 53 min |
Heat emission at 250 ?C
| 760 W |
Air exchange speed at 250 ?C
| 49 cca/hour |
Electricity – mains 50/60 Hz
| max. power input | 1,9 kW |
| nominal voltage | 230 V |
| current | 8,3 A |
| stand by | 5 W |
























