Laboratory Drying Ovens VENTICELL 707

PRODUCT OVERVIEW

VENTICELL 707

Ovens with Forced Circulation
-Type Eco Line MC000215 -Type Evo Line MC001819 Due to a patented ventilation system the air within the VENTICELL® chamber is ventilated in a regular spiral way. This leads to a homogenous temperature profile throughout the chamber and short heating times. Operating economy is ensured by higher rate and precision of heating in laboratories. Especially suitable for very moist goods.
  • precise and fast laboratory drying oven
  • chamber volume 707 litres
  • temperature range from 10°C above ambient to 250/300°C
  • forced air circulation system in the chamber guarantees homogeneous temperature distribution in all the processes of drying, heating and sterilization of materials in laboratories
  • high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
  • higher speed and accuracy of all the tempering procedures ensures economical operation, especially suitable for high humidity materials, for demanding and accurate tests and processes
  • wide selection of optional equipment and accessories
  • output for connecting a printer or PC
  • possibility of validation (IQ, OQ)

Specification

Interior of stainless steel material No. 1.4301

volume

707 l

depth

520 mm

width

940 mm

height

1410 mm

Tray

standard equipment

2 pcs

storage area (w × d)

920 × 485 mm

racks

19 max. number

distance between guide rails

70 mm

Maximal weight of the load *)

inside the oven

130 kg/oven

per tray

50 kg/tray

Number of doors

2 pcs

External dimensions (including door and handle)

depth

790 mm

diameter of the air branch outer/inner

52/49 mm

width

1160 mm

height (inclusive foots – N)

1910 (K) mm

Package dimensions (three layers carton)

depth

860 mm

width

1230 mm

depth

2085 mm

Weight

brut

240 kg

net

215 kg

Working temperature **) (beginning of the regulation) – VENTICELL / VENTICELL plus

from 10 °C above ambient temp. to

250 °C

Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door, > 50 °C

time variation cca (±)

0,4 °C

space deviation (±) % cca (±)% of the reached temperature

2,5 %

Time required to reach 250 °C with closed air flap and voltage 230 V

64 min

Heat emission at 250 °C

2550 W

Air exchange speed at 250 °C

12 cca/hour

Electricity – mains 50/60 Hz

max. power input

4,9 kW

nominal voltage

400/3N V

current

7,8 A

stand by

5 W

 