TurboQ Real-Time PCR System
Ideal for Multiplex and Routine PCR Applications
TurboQ Real-Time PCR system is designed for multiplex and routine PCR applications with up to 4 targets. Inheriting the proven PCR instrument core competency from Blue-Ray Biotech, TurboQ brings versatile features to help lab researchers in performing diagnostic testing and researches.
Features and Benefits
![4-optical-channels](https://www.blue-raybio.com/uploads/files/4-optical-channels.png)
4 Optical Channels
Detects up to 4 targets in one well![ccd-based-detection-system](https://www.blue-raybio.com/uploads/files/ccd-based-detection-system.png)
CCD-Based Detection System
The entire plate can be detected at once to eliminate reaction-time differences![motorized-plate-carrier-tray](https://www.blue-raybio.com/uploads/files/motorized-plate-carrier-tray.png)
Motorized Plate Carrier Tray
Plate in/out with a simple click, enables automation integration in the future![0.25°c-block-temperature-control-accuracy](https://www.blue-raybio.com/uploads/files/block-temperature-control-accuracy.png)
+/- 0.25°C Block Temperature Control Accuracy
Ensures experiment condition stability and eliminates uncertainty![software-with-complete-analysis-functions](https://www.blue-raybio.com/uploads/files/software-with-complete-analysis-functions.png)
Software with Complete Analysis Functions
Functions include Absolute Quantitation, Relative Quantitation, Allelic Discrimination, Melting Curve. No need to purchase extra analysis modulesSpecifications
Sample Block
- Sample Block
- Fixed 96-well, compatible with regular profile 0.2 ml PCR tube, strip, and semi-skirted 96-well plate
- Sample Volume
- 10 – 100 µl
- Plate Loading
- Motorized plate carrier tray
Block Temperature
- Block Temperature Range
- 25 – 99 °C
- Peak Block Ramp Rate
- 2.5 °C/sec
- Temperature Accuracy
- ± 0.25 °C (35 °C – 95 °C) of set point/ display temperature / Measured at 3 minutes after clock start.
- Temperature Uniformity Across Block
- ± 0.5 °C (30 seconds after clock start)
Gradient Temperature
- Gradient Temperature Range
- 60 – 99 °C
- Gradient Temperature Difference
- Max. span 25 °C
Heater Lid
- Heating Lid
- 105°C
Optics
- Excitation Source
- Tungsten-halogen lamp
- Detection Module
- CCD camera
- Detection Channel
- 4 channels (Ch1: FAM, SYBR Green, Ch2: HEX, VIC, JOE, Ch3: ROX, Texas Red, Ch4:Cy5)
Environment
- Operating Temperature
- 15 – 30 °C
- Operating Humidity
- RH: 85% or less
- Storage Temperature
- -10 – 60 °C
General
- PC Operation System
- Windows 7 or higher
- Footprint Dimensions (H x W x D)
- 567 mm x 354 mm x 455 mm
- Weight
- 35 kg
- Power Supply
- 110/230 V, frequency 50/60 Hz, 1500W
- Fuse
- 1 x fuse, 10A for 110VAC power / 5A for 220VAC power