Mainly used in Lab applications. Excellent EMC performance, ESD≥Level IV (IEC 61000-4-2). The acceptable syringe specification is from 10 μL to 60 mL. Suitable for high accuracy and small flow rate liquid transferring.
Flow rates:0.001μl-43.349ml/min
Max. Number of Channels:1
Product Overview
Applications:
Mainly used in Lab applications. Excellent EMC performance, ESD≥Level IV (IEC 61000-4-2)
Product Instruction:
The acceptable syringe specification is from 10 μL to 60 mL. Suitable for high accuracy and small flow rate liquid transferring.
Specifications
Main Functions and Features
– Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
– Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt.
– Working Mode: Infusion
– Memory Function:
1. The parameters are saved in EEPROM. The parameters don’t need to be reset when power returns after an interruption
2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns after an interruption
– Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked
– Communication Function: Realize computer control through RS485 communication interface
– External Control Function: Input/output control
– Calibration Function: Acquire accurate volume through calibration
– Syringe Protection: Adjust syringe rest to prevent syringe from damaging
Technical Specifications
Syringe pump:LSP01-3A
No. of Syringes:1
Advance per Mircostep:1/16step: 0.156 μm
Infusion Volumn per Microstep:0.13 μL (60 mL BD Syringe)
Syringe Size:10 μL – 60 mL
Max. Linear Rate:65 mm/min
Min. Linear Rate:5μm/min
Flow Rate:0.831 nL/min – 54.155 ml/min
Working Mode:Infusion
Linear Force:> 9 kgf
Max. Step Rate:6933 (1/16step)/ sec
Min. Step Rate:16 (1/16 step) /30 sec
Accuracy:≤ 0.5% error in the condition of > 30% of max. infusion distance
Setting Mode:Rotary coded switch and membrane keypad
Display:128 x 64 graphic LCD
Power:AC 196V-240V/20W
Operating Condition:Temperature 5 °C – 40 °C Relative humidity < 80 %
Dimensions:280 × 210 × 140 mm
Weights:3.6 kg