Incubator INCUCELL V 404

PRICE UPON REQUEST

INCUCELL V 404

Incubators with Forced Circulation

-Type ECO Line MC000231

-Type EVO Line MC000701

Suitable for safe processing of microbiological cultures. The INCUCELL® V line (with a fan) ensures quicker and more precise temperature distribution with small deviations. These units are to be used especially in biological and microbiological laboratories, quality tests in the pharmacy, cosmetics and testing in the veterinary medicine and food processing industry.
  • natural, accurate – laboratory incubator
  • chamber volume 404 litres
  • temperature range from 10°C above ambient to 100°C
  • characterized by noiseless operation and fine air flow in the usable space
  • high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
  • high-quality and safe drying, heating and tempering of common laboratory materials in laboratory and industrial areas, suitable for safe processing of microbiological cultures
  • standard version with microprocessor control system Fuzzy logic, controls placed on a foil keyboard, process information displayed on LED display
  • wide selection of optional equipment and accessories
  • output for connecting a printer or PC
  • possibility of validation (IQ, OQ)

Interior of stainless steel material No. 1.4301

volume

404 l

width

540 mm

height

1410 mm

Tray

standard equipment

2 pcs

storage area (w × d)

520 × 485 mm

racks

19 max. number

distance between guide rails

70 mm

Maximal weight of the load *)

inside the oven

100 kg/oven

per tray

30 kg/tray

Door

1 pcs

External dimensions (including door and handle)

depth

790 mm

diameter of the air branch outer/inner

52/49 mm

width

760 mm

1910 (K)

Package dimensions (three layers carton)

depth

860 mm

width

830 mm

depth

2085 mm

Weight

brut

175 kg

nett

150 kg

Working temperature (beginning of the regulation)

from 10 °C above ambient temp. to:

100 °C

Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door

time variation cca (±)

<=0,2 °C

space deviation cca (±)

<=0,8 °C

Time required to reach 37 °C with closed air flap and voltage 230 V

38 min

Heat emission at 37 °C

115 W

Air exchange speed at 37 °C

18 per hour

Electricity – mains 50/60 Hz

max. power

1,3 kW

nominal voltage

230 V

current

5,6 A

power by stand by

5 W