INCUCELL V 404
Incubators with Forced Circulation
-Type ECO Line MC000231
-Type EVO Line MC000701
Suitable for safe processing of microbiological cultures. The INCUCELL® V line (with a fan) ensures quicker and more precise temperature distribution with small deviations. These units are to be used especially in biological and microbiological laboratories, quality tests in the pharmacy, cosmetics and testing in the veterinary medicine and food processing industry.- natural, accurate – laboratory incubator
- chamber volume 404 litres
- temperature range from 10°C above ambient to 100°C
- characterized by noiseless operation and fine air flow in the usable space
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- high-quality and safe drying, heating and tempering of common laboratory materials in laboratory and industrial areas, suitable for safe processing of microbiological cultures
- standard version with microprocessor control system Fuzzy logic, controls placed on a foil keyboard, process information displayed on LED display
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Interior of stainless steel material No. 1.4301
volume | 404 l |
width | 540 mm |
height | 1410 mm |
Tray
standard equipment | 2 pcs |
storage area (w × d) | 520 × 485 mm |
racks | 19 max. number |
distance between guide rails | 70 mm |
Maximal weight of the load *)
inside the oven | 100 kg/oven |
per tray | 30 kg/tray |
Door
1 pcs |
External dimensions (including door and handle)
depth | 790 mm |
diameter of the air branch outer/inner | 52/49 mm |
width | 760 mm |
1910 (K) |
Package dimensions (three layers carton)
depth | 860 mm |
width | 830 mm |
depth | 2085 mm |
Weight
brut | 175 kg |
nett | 150 kg |
Working temperature (beginning of the regulation)
from 10 °C above ambient temp. to: | 100 °C |
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
time variation cca (±) | <=0,2 °C |
space deviation cca (±) | <=0,8 °C |
Time required to reach 37 °C with closed air flap and voltage 230 V
38 min |
Heat emission at 37 °C
115 W |
Air exchange speed at 37 °C
18 per hour |
Electricity – mains 50/60 Hz
max. power | 1,3 kW |
nominal voltage | 230 V |
current | 5,6 A |
power by stand by | 5 W |