Incubator INCUCELL V 55

PRICE UPON REQUEST

INCUCELL V 55

Incubators with Forced Circulation
-Type ECO Line MC000228 -Type EVO Line MC000684 Suitable for safe processing of microbiological cultures. The INCUCELL® V line (with a fan) ensures quicker and more precise temperature distribution with small deviations. These units are to be used especially in biological and microbiological laboratories, quality tests in the pharmacy, cosmetics and testing in the veterinary medicine and food processing industry.
  • natural, accurate – laboratory incubator
  • chamber volume 55 litres
  • temperature range from 10°C above ambient to 100°C
  • characterized by noiseless operation and fine air flow in the usable space
  • high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
  • high-quality and safe drying, heating and tempering of common laboratory materials in laboratory and industrial areas, suitable for safe processing of microbiological cultures
  • standard version with microprocessor control system Fuzzy logic, controls placed on a foil keyboard, process information displayed on LED display
  • wide selection of optional equipment and accessories
  • output for connecting a printer or PC
  • possibility of validation (IQ, OQ)

Specification

Inner dimensions stainless steel DIN 1.4310

volume

55L

Broad

400mm

Height

350mm

shelf supports

Standard Delivery

2 pieces

Storage area (w × d)

380×335mm

Guides for the shelves

4 maximum number

Max. permissible load on the shelves

70mm

Max. permissible load on the shelves *)

total carrying capacity

50kg/cabinet

load capacity of the trays

20 kg/tray

Number of doors

1 piece

External dimensions (incl. doors and handle)

depth

680mm

Outer/inner vent diameter

52/49mm

Broad

620mm

680 (N)

Packaging – dimensions – three-layer cardboard

Height

760mm

Broad

700mm

Height (including the pallet)

880mm

Weight

gross

66kg

net

55kg

Working temperature (beginning of regulation)

of 10 °C above the ambient temp. until:

100ºC

Temperature deviations according to DIN 12 880 Part 2, from the working temp. With the air damper and door closed

Temporal ca (±)

<=0.2°C

Spatially cca ±

<=0.3°C

Start-up time to the temperature of 37 °C with the air damper closed and a voltage of 230 V

41 mins

Heat dissipation to ambient at 37°C

80W

Air exchange rate at 37 °C

45 in hour

Electrical parameters – mains 50/60 Hz

Max. Input

0.7kW

Tension

230V

Electricity

3 A

Power consumption in the ready position (stand by)

5w