INNO-H (Absorbance, Luminescence & Fluorescence Microplate Reader)

SKU: N/A Categories: , Brand:
General
Detection modes UV-Vis Absorbance
Fluorescence Intensity (Monochromator)
Dichroic Intensity (Filter)
Fluorescence Polarization (Filter)
TRF (Time-Resolved Fluorescence)
Luminescence (Monochromator/Filter)
Read methods End point
Kinetic
Spectral scanning
Well-area scanning
Microplate types 6-to 384-well plates
Others labware supported NANO-VC (Microvolume Plate)
INNO-QM (Quality Check Plate)
Temperature control Up to 45 ?C ? 0.2?c at 37?c
Shaking Linear & Orbital & Double Orbital with 4 different speeds
Software INNO-XTM?(Basic software) & INNO-XSTM?(21 CFR part 11 Compliance Software) (Optional)
Certifications CE Marked
RoHS
ISO 9001 / ISO 13485 / ISO 14001
Absorbance
Light source Xenon flash lamp
Detector Photodiode
Wavelength selection Monochromator
Wavelength range 230?999 nm, 1nm increments
Dynamic range 0?4.0 OD
Resolution 0.0001 OD
Pathlength correction Yes
Wavelength accuracy ?2 nm
Wavelength repeatability ?0.2 nm
OD linearity <1% from 0 to 3.0 OD
OD repeatability < 0.5% at 2.0 OD
Fluorescence Intensity (Monochromator-based Fluorescence)
Light source Xenon flash lamp
Detector PMT
Wavelength selection Monochromator (Bandwidth Variable Option)
Wavelength range 250~700 nm (Options 850nm)
Dynamic range >7 decades
Sensitivity Fluorescein 2.5 pM top / 5pM bottom (96well plate)
Dichroic Intensity (Filter-based Fluorescence)
Light source Xenon flash lamp
Detector PMT
Wavelength selection Filters
Wavelength range 250~700 nm (Options 850nm)
Dynamic range >7 decades
Sensitivity Top Fluorescein 0.25 pM (96-well plate)
Fluorescence Polarization (Filter-based Fluorescence)
Light source Xenon flash lamp
Detector PMT
Wavelength selection Filters
Wavelength range 400~700 nm
Sensitivity 2 mP at 1nM fluorescein
Luminescence
Detector PMT
Wavelength selection Filters / Monochromator
Wavelength Range 200~700 nm (Options 850nm)
Sensitivity 10 amol ATP (filter) 20 amol ATP (monochromator)
TRF (Time-Resolved Fluorescence)
Light source Xenon flash lamp
Detector Filters / Monochromator
Wavelength selection PMT
Physical Characteristics
Connectivity 1 USB, 1 RS232 for external PC control
Power 100-240 Volts AC. 50/60 Hz
Dimensions 500W x 410L x 360H
Weight 25 kg
Applicable optional products
Reagent Injector (INNO-D)
  • 2 Syringe pumps
  • 5~1,000? ? 15~1,000?
  • Minimum prime Vol. 1.1mL, 100?L with back flush
Microvolume Plate (NANO-VC)
  • Nucleic acid and protein quantification with DNA/RNA, Lysozyme, dsDNA, ssDNA, Etc.
  • 2-2.5?L sample capacity with 24 wells
  • 2.5ml cuvette holder
Quality Check Plate (INNO-Q)
  • Absorbance linearity, accuracy, and alignment QC
INNO-H product line configuration chart
The user can select two filters by default when purchasing the INNO-H(D)
INNO-H INNO-HML INNO-HM INNO-HD
Monochromator Absorbance ? ? ?
Monochromator Fluorescence ? ? ?
Monochromator Luminescence (Fiber) ? ?
Filter Fluorescence (Dichroic) ? ?
Filter Luminescence ? ?
Fluorescence Polarization (Option) ? ?
TRF ? ? ? ?
Shaking & Incubation ? ? ? ?
INNO product line series configuration chart
INNO INNO-M INNO-SA INNO-SF INNO-S INNO-HD INNO-HM INNO-HML INNO-H
Absorbance (Monochromator) O O O O O O O
Luminescence (Filter) O O O O O
Luminescence (Monochromator) O O
Fluorescence (Filter) O O O O
Fluorescence (Monochromator) O O O
TRF O O O O O O
Dichroic Intensity O O
Fluorescence Polarization O O
Shaking O O O O O O O O O
Incubation O O O O O O O
Dual Injector O O O O O O O
Fluorescence Dichroic Filter List (INNO-H&HD)
The user can select two filters by default when purchasing the INNO-H(D)
NO Filter NAME
1 CFP 11 DAPI
2 GFP 12 Lucifer Yellow
3 FITC 13 SpectrumGreen
4 YFP 14 Qdot525
5 CY3 15 Qdot605
6 TRITC 16 Qdot625
7 Texas Red 17 Qdot655
8 mCherry 18 SpectrumGold
9 CY5 19 SpectrumOrange
10 CY5.5 20 SpectrumRed

 

DOWNLOAD BROCHURE