FRIOCELL 111
Laboratory Incubators with Forced Air Circulation with/for Heating and Cooling
-Type ECO Line MC000235
-Type EVO LineÂ
The high technical standard of our FRIOCELL® incubators allows exact incubation processes both for variation and deviation. The units have very short recovery times and show an excellent results in keeping the precise regulation. A unique cooling system ensures, that the samples are not dried while cooling. A high Âperformance system of lighting ensures outstanding homogenous parameters for tests and growth conditions. These devices are desigÂned for use in biotechnology, botany, food processÂing industry, cosmetics, chemical industry etc.- precise, adaptable to nature, economical – laboratory incubator
- chamber volume 111 litres
- temperature range: 0°C to 100°C
- patented cooling system (ecological working medium R134a) offers accurate and fast simulation of natural processes and reduces the risk of evaporation of samples, the devices may be completed with various types of inner chamber lighting
- high operation comfort, precise temperature control and the possibility of achieving extremely short times of temperature changes as well as short times of temperature equalization in the chamber after the door opening; standard equipment also includes inner glass door and possibility of fan speed regulation
- suitable for demanding and accurate specimen culture testing and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Specification
Inner dimensions – Chamber, stainless steel
volume | 111 l |
depth | 370 mm |
width | 540 mm |
height | 530 mm |
Volume of the steam space
163 cca l |
Trays, stainless steel *)
Min. distance between trays | 70 mm |
number | 7/2 max./usual |
Storage area (w × d) | 520 × 335 mm |
Number of outer metal doors
number | 1 pcs |
Number of inner glass doors
number | 1 pcs |
Admissible weight of trays
together inside the oven | 50 kg |
per 1 tray | 20 kg |
Electricity data
Max. power | 700 W |
IP 20 | |
mains 50/60Hz | 230 V |
Temperature data
Cooling/down time from 22 °C to 10 °C | < 21 min |
Heating/up time to 37 °C from the ambient temperature | 11 min |
Working temperature from 0,0 to | 100 °C |
Temperature accuracy – in time | 0,2 ± °C |
Temperature accuracy – in space at 10 °C | 0,5 ± °C |
Recovery time after 30 s door open
at 37 °C | 5 min |
at 50 °C | 6 min |
Heat emission
at 37 °C | 70 W |
Outer dimensions, incl. door and handle, Feet (N) or Rollers (K)
depth | 750 mm |
width | 760 mm |
height | 1045 (N) mm |
Weight
brut | 136 kg |
nett | 106 kg |
Â
Â